WALL MOUNTING INSTRUCTIONS
MAINTENANCE
CONNECTING THE INFLATION
SYSTEM TO MANOMETER
Remove red safety cap which seals the
reservoir containing mercury (on the
952, the red safety cap is located on the
air inlet beneath swivel base). Attach
free end of reservoir tube securely to air
inlet. Save red safety cap for later use.
When transporting mercury instrument
we recommend placing the red safety
cap on reservoir.
Please Note: It is normal for negligible
amounts of residual mercury droplets
to accumulate around the air inlet
beneath the red safety cap during trans-
portation.
To Lock Mercury Within the Reservoir: Tilt the entire unit back
45° towards the reservoir to permit mercury to flow out of
cartridge tube and into reservoir. When cartridge tube is
completely emptied of ALL mercury (and while it is still tilted
45°), move locking lever to the right “OFF”. Mercury should be
locked within the reservoir during maintenance or transport. To
release mercury, move lever to “ON” position. If necessary, tilt
unit towards tube to allow mercury to flow out of reservoir.
Lift on the spring loaded EZ-Tube
mechanism
located at the top of the instrument. Tilt instrument
forward to free tube. If necessary use a small
flathead screwdriver to help dislodge the top of the
tube. To replace tube, lift EZ-Tube
mechanism and
insert base of tube into main unit, being sure
graduation marks face out. Next, position top of
tube and release EZ-Tube
mechanism.
®
To Register Your Product, visit us at www.adctoday.com and follow the links
ADC
55 Commerce Drive
Hauppauge, NY 11788
ADC (UK) Ltd.
Unit 6, PO14 1TH
United Kingdom
tel: 631-273-9600, 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: info@adctoday.com
EC REP
IB p/n 9354A-00 - rev 1 ©2007 ADC • Printed in U.S.A.
Tools Needed
Medium slotted screwdriver
Hammer
Drill with 3/16” drill bit
Wall Mount Instructions
1. Unit may be wall mounted without disassembling main unit from swivel
base/wall mount bracket. Position bracket so that center of unit is
approximately 4’6” from the floor. Using a pencil, mark the holes on either
side of the wall bracket.
2. Check for level and drill 3/16” holes approximately 7/8” deep. Gently tap the 2
screw anchors into the holes until they are flush with the wall.
3. Position the manometer and bracket and mount to the wall using supplied
screws. Take care not to overtighten screws. Position the wall basket as
shown in the illustration at right, approximately 2” to the right of the
manometer.
4. Using a pencil, mark the 3 holes as before. Repeat the procedure for drilling
the holes, inserting anchors, and mounting.
5. Attach 8 foot length of coiled tubing to air inlet underneath mounting bracket.
Insert “male” luer connector on free end of coiled tubing to “female” luer
connector on bladder tubing. Wrap cuff around bulb and store in the basket.
Important: This instrument is designed to permit wall mounting without first
disassembling main unit from its swivel base/mounting bracket.
To remove main unit from bracket, disconnect tubing from mounting bracket.
Unscrew wingnut from bottom of mounting bracket, separating main unit from
bracket. Reverse procedure to reassemble.